  • Inductively coupled plasma mass spectrometer (ICPMS)

    Author: Release time:2021-06-01 Views:1675


    Introduction: The inductively coupled plasma mass spectrometer is the instrument of measuring the trace content of elements and ratio of isotopes. The advantage of this instrument is that it has low detection limit (ng/ml or lower), the base effect is small,simple line, and could determine many elements at the same time. On the other hand, dynamic linear range is wide and could determine the isotope ratios rapidly.

    Instrument name

    Inductively coupled plasma mass spectrometer (ICPMS)

    Model

     M90

    Manufacturer  

    analytikjena

    Secifications

    1. Sensitivity High sensitivity mode:

    Low mass number Be (9) > 100 M cps/ppm;

    Medium mass number In(115)> 1000M cps/ppm;

    High mass number Th(232) > 500 M cps/ppm;

    2. Oxide ion ration

    CeO+ / Ce+ < 2 %(Normal sensitivity mode)

    CeO+ / Ce+ < 3 %(High sensitivity mode)

    3. Bivalent ion ration: Ba++ / Ba+ < 3 %

    4. background(5amu)< 2 c/s

    5. Short-time precision (20 min)< 3 %

    6. Long-time stability(4 hours)< 4 %

    7. abundance sensitivity low-end 5*10-7, high-end 1*10-7

    8. accuracy of isotope ratio< 0.1 % (107Ag/109Ag)

    9. element detection limit(ng/L) 9Be<0.5; 115In<0.1; 232Th<0.05

    Applications and Uses

    ICPMS is frequently used to analyze various elements.

    Accessories

    None